柔性导电复材电极片在动态心电图适配采集中的应用

  • 打印
  • 收藏
收藏成功


打开文本图片集

中图分类号:R318.08;TQ342+.87 文献标志码:A 文章编号:1001-5922(2026)03-0729-04

Application of flexible conductive composite electrode sheets in adaptive acquisition of dynamic electrocardiogram

GUO Meiqin

(Electrocardiogram(ECG) Department Wuxue First People's Hospital, Wuxue 435499,Hubei China)

Abstract:The traditional dynamic electrocardiogram electrode shets feature poor substrate flexibility and uneven dispersion of conductive filers,which make them diffcult toadapt to the loose skinoftheelderly.This easily causes interfacial gapsand motionartifacts,leading tounstable signals.To improve theelectrical conductivityandanti-interference performance offlexibleconductivecomposite materials foradaptation tothe elderly'sskin,this studyprepared flexible conductivecomposite electrode sheets by designing four experimental groups and one control group,so as to monitorthe changes in interfacial contact impedanceand anti-interferenceperformanceagainst motion artifacts.The experimental results showed that Experimental Group 3( 0.600kg of polyurethane adhesive, 0.030kg of carbon black , 0.010kg of multi-walled carbon nanotubes , 0.007kg of graphene, 0.015kg of polyethylene glycol) was the optimal group,with an average impedance of 25. 1kΩ ,a decrease of 61.1% compared with the control group.In addition,the signal distortion rate of Experimental Group 3 was 7.8% ,which was lower than that of all other groups.

Key words:flexible conductive composite materials;geriatrics;dynamic electrocardiogram; electrode shets;signal stability

柔性导电复合材料作为兼具优异导电性与良好柔韧性的新型功能材料,近年来在生物医疗领域展现出广阔的应用前景。(剩余5208字)